Detailed profile:
User manual.pdf
LC98II Liquid Chromatography
P98II Isocratic Pump
1. Specification a. Operating pressure: 0~42MPa (6000 psi) maximum. b. Flow rate range: 0.01~9.99ml/min c. Flow rate precision: ±1% d. Flow rate accuracy: ±3% e. Pressure solution: 0.1MPa f. Over pressure protection: pump shuts off if pressure exceeds limit specified by user or 42.0MPa(6000 psi) maximum. 2. General Parameters Power Supply: 200W Dimensions: 450mm(D)×260mm(W)×150mm(H) Weight: 15kg
UV98II UV Detector 1. Specification Wavelength range: 190~700nm Spectral bandwidth: 5nm Wavelength accuracy: ±2nm Wavelength reproducibility: ±0.2nm Absorbance range: Recorder output: 0.002au/mV to 2au/mV in steps of 1, 2, 5 Data system output: 2au/V, fixed Noise: 5×10-5au (Peak-Peak) Test Conditions: Wavelength: 254nm Mobile phase: methanol Flow rate: 1ml/min Time constants: 0.5s
Drift: 1×10-3au/hr Test Conditions: Wavelength: 254nm Mobile phase: methanol Flow rate: 1ml/min Time constants: 0.5s Flow cell volume: 4.5μl Time constants: 0.05s, 0.5s, 5s Autozero: Yes 2. General Parameters Power Supply: 200W Dimensions: 450mm(D)×260mm(W)×150mm(H) Weight: 15kg
T98 Column Oven 1. Specification Temperature range: ambient ~150C. Resolution: 1C steps Precision: ±0.5C 2. General Parameters Power Supply: 100W Dimensions: 450mm(D)×260mm(W)×150mm(H) Weight: 5kg
|