Product
  • Instrument Analytical instrument Biology instrument Healthcare instrument Histology instrument Hospital & Lab instrument Other instrument Consumable ESU pencil Histology Consumable Microtome blade holder Slides and Cover Glass Slides storage
  • Product details
    NameLC98I
    Detailed profile:

                                                                               User manual.pdf

     

    P98I Binary Gradient Pump

    1. SPECIFICATION

    a. Operating pressure: 0-42MPa (6000 psi) maximum.

    b. Flow rate range: 0.1-15ml/min,

    c. Flow rate precision: 1%,

    d. Flow rate accuracy: 25ml/368-387s(3%),

    e. Compositional precision:0.5% absolute

    f. Pressure solution: 0.1MPa,

    g. Over pressure protection: pump shuts off if pressure exceeds limit specified by user-42.0 MPa(6000 psi) maximum.

    2. Environment

    a.Temperature requirements: ambient temperature range 10 to 35 degree

    b. Operates in relative humidity from 5% to 80%;

    c. Power requirements: 220Vac 5%,

    d.Power frequency: 50 0.5Hz (voltage and frequency specified on the tag attached to the power cord)

    3.  GENERAL PARAMETERS

    a. Power Supply: 400W,

    b. Dimensions: 450mm(D)×210mm(W)×390mm(H),

    c. Weight: 20kg,

    UV98I UV Detector
    1. Specification
    Wavelength range: 190-700nm

    Spectral bandwidth:5nm

    Wavelength accuracy: 2nm

    Wavelength reproducibility: 0.2nm

    Absorbance range:

    Recorder output: 0.002au/mV to 2au/mV in steps of 1, 2, 5

    Data system output: 2au/V, fixed

    Noise: 5×10-5au (Peak-Peak)

    Test Conditions:  Wavelength: 254nm

    Mobile phase: methanol

    Flow rate: 1ml/min

    Time constants: 0.5s

    Drift: 1×10-3au/hr

    Test Conditions: Wavelength: 254nm

    Mobile phase: methanol

    Flow rate: 1ml/min

    Time constants: 0.5s

    Flow cell volume:4.5ul

    Time constants: 0.05s, 0.5s, 5s

    Autozero:Yes

    2.  General Parameters

    a. Power Supply: 200W,

    b. Dimensions: 450mm(D)x180mm(W)×390mm(H),

    c. Weight: 15kg

     

    T98 Column Oven

    Temperature controller and column oven combined in one or as individual unit.

    a.  temperature range: ambient-150

    b.  precision: ±0.5

    c.  capacity of column oven: combined unit 300mm, long column 4 pieces.

     

    N2000 Work Station

    Double channel, Modem, GMP certificate.

     

    Rheodyne 7725i Injector

    Pressure up to 42MPa, nice feeling.

    【Nothing】 【Next
    Add: No. 5 Hepanxiang, Jinhua, Zhejiang 321001 China      E-mail: info@mossinstruments.com
    Copyright© Moss Instruments Co., Ltd    Design: jhhuosu